Preliminary research on dose response of LiF thermoluminescence dosimeter to irradiation of 49.3 MeV/u carbon ion beam
文献类型:期刊论文
作者 | Li Qiang![]() ![]() ![]() ![]() ![]() |
刊名 | Nuclear Techniques
![]() |
出版日期 | 1999 |
卷号 | 22页码:636-9 |
ISSN号 | 0253-3219 |
英文摘要 | LiF(Mg,Ti) thermoluminescence dosimeter (TLD) was applied to measure dose response to irradiation of 49.3 MeV/u 12C ion beam supplied by HIRFL. Moderate variance of dose-response efficiency was found when particle fluence increases from 1times107 to 2times108/cm2. The average dose-response efficiency 57% was obtained. This phenomenon implies that LiF can be used to measure absorbed dose in a certain range of particle fluence. Moreover, the result is in a good agreement with the calculated dose-response efficiency 54% to 50 MeV/u 12C ion beam, which was predicted by the model of heavy-ion track structure at GSI, Germany. |
WOS记录号 | INSPEC:6538633 |
源URL | [http://119.78.100.186/handle/113462/59921] ![]() |
专题 | 近代物理研究所_兰州重离子研究装置 近代物理研究所_生物物理研究室 |
推荐引用方式 GB/T 7714 | Li Qiang,Wang Guiling,Wei Zengquan,et al. Preliminary research on dose response of LiF thermoluminescence dosimeter to irradiation of 49.3 MeV/u carbon ion beam[J]. Nuclear Techniques,1999,22:636-9. |
APA | Li Qiang.,Wang Guiling.,Wei Zengquan.,Li Wenjian.,Dang Bingrong.,...&Zhou Guangming.(1999).Preliminary research on dose response of LiF thermoluminescence dosimeter to irradiation of 49.3 MeV/u carbon ion beam.Nuclear Techniques,22,636-9. |
MLA | Li Qiang,et al."Preliminary research on dose response of LiF thermoluminescence dosimeter to irradiation of 49.3 MeV/u carbon ion beam".Nuclear Techniques 22(1999):636-9. |
入库方式: OAI收割
来源:近代物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。