中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Potentiometric flow injection system for determination of reductants using a polymeric membrane permanganate ion-selective electrode based on current-controlled reagent delivery

文献类型:期刊论文

作者Song, Wenjing1,2; Ding, Jiawang1; Liang, Rongning1; Qin, Wei1
刊名ANALYTICA CHIMICA ACTA
出版日期2011-10-17
卷号704期号:1-2页码:68-72
关键词Potentiometry Flow injection analysis Ion-selective electrodes Reductants Current-controlled release
产权排序Chinese Acad Sci, Shandong Prov Key Lab Coastal Zone Environm Proc, Yantai Inst Coastal Zone Res, Yantai 264003, Peoples R China;Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
通讯作者Qin, W (通讯作者),Chinese Acad Sci, Shandong Prov Key Lab Coastal Zone Environm Proc, Yantai Inst Coastal Zone Res, Yantai 264003, Peoples R China
中文摘要A polymeric membrane permanganate-selective electrode has been developed as a current-controlled reagent release system for potentiometric detection of reductants in flow injection analysis. By applying an external current, diffusion of permanganate ions across the polymeric membrane can be controlled precisely. The permanganate ions released at the sample-membrane interface from the inner filling solution of the electrode are consumed by reaction with a reductant in the sample solution thus changing the measured membrane potential, by which the reductant can be sensed potentiometrically.Ascorbate, dopamine and norepinephrine have been employed as the model reductants. Under the optimized conditions, the potential peak heights are proportional to the reductant concentrations in the ranges of 1.0 × 10−5 to 2.5 × 10−7 M for ascorbate, of 1.0 × 10−5 to 5.0 × 10−7 M for dopamine, and of 1.0 × 10−5 to 5.0 × 10−7 M for norepinephrine, respectively with the corresponding detection limits of 7.8 × 10−8, 1.0 × 10−7 and 1.0 × 10−7 M. The proposed system has been successfully applied to the determination of reductants in pharmaceutical preparations and vegetables, and the results agree well with those of iodimetric analysis.
英文摘要A polymeric membrane permanganate-selective electrode has been developed as a current-controlled reagent release system for potentiometric detection of reductants in flow injection analysis. By applying an external current, diffusion of permanganate ions across the polymeric membrane can be controlled precisely. The permanganate ions released at the sample-membrane interface from the inner filling solution of the electrode are consumed by reaction with a reductant in the sample solution thus changing the measured membrane potential, by which the reductant can be sensed potentiometrically. Ascotbate, dopamine and norepinephrine have been employed as the model reductants. Under the optimized conditions, the potential peak heights are proportional to the reductant concentrations in the ranges of 1.0 x 10(-5) to 2.5 x 10(-7) M for ascorbate, of 1.0 x 10(-5) to 5.0 x 10(-7) M for dopamine, and of 1.0 x 10(-5) to 5.0 x 10(-7) M for norepinephrine, respectively with the corresponding detection limits of 7.8 x 10(-8), 1.0 x 10(-7) and 1.0 x 10(-7) M. The proposed system has been successfully applied to the determination of reductants in pharmaceutical preparations and vegetables, and the results agree well with those of iodimetric analysis. (C) 2011 Elsevier B.V. All rights reserved.
学科主题化学
研究领域[WOS]Chemistry
关键词[WOS]L-ASCORBIC-ACID ; AMINO-ACIDS ; COPPER WIRE ; FIA SYSTEM ; DETECTOR ; DOPAMINE ; ANIONS
收录类别SCI
资助信息Outstanding Youth Natural Science Foundation of Shandong Province[JQ200814]; Chinese Academy of Sciences; National Natural Science Foundation of China[20977073]; Taishan Scholar Program of Shandong Province
语种英语
WOS记录号WOS:000296824600007
源URL[http://ir.yic.ac.cn/handle/133337/5166]  
专题烟台海岸带研究所_中科院海岸带环境过程与生态修复重点实验室
烟台海岸带研究所_近岸生态与环境实验室
作者单位1.Chinese Acad Sci, Shandong Prov Key Lab Coastal Zone Environm Proc, Yantai Inst Coastal Zone Res, Yantai 264003, Peoples R China
2.Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
推荐引用方式
GB/T 7714
Song, Wenjing,Ding, Jiawang,Liang, Rongning,et al. Potentiometric flow injection system for determination of reductants using a polymeric membrane permanganate ion-selective electrode based on current-controlled reagent delivery[J]. ANALYTICA CHIMICA ACTA,2011,704(1-2):68-72.
APA Song, Wenjing,Ding, Jiawang,Liang, Rongning,&Qin, Wei.(2011).Potentiometric flow injection system for determination of reductants using a polymeric membrane permanganate ion-selective electrode based on current-controlled reagent delivery.ANALYTICA CHIMICA ACTA,704(1-2),68-72.
MLA Song, Wenjing,et al."Potentiometric flow injection system for determination of reductants using a polymeric membrane permanganate ion-selective electrode based on current-controlled reagent delivery".ANALYTICA CHIMICA ACTA 704.1-2(2011):68-72.

入库方式: OAI收割

来源:烟台海岸带研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。