Influences of the thickness, misalignment, and dispersion of the Savart polariscope on the optical path difference and spectral resolution in the polarization interference imaging spectrometer
文献类型:期刊论文
作者 | Zhang, Chunmin1; Zhao, Jianke1,2; Sun, Yao1,3 |
刊名 | applied optics
![]() |
出版日期 | 2011-07-10 |
卷号 | 50期号:20页码:3497-3504 |
关键词 | UPPER ATMOSPHERIC WIND FOURIER-TRANSFORM ORIENTATION |
ISSN号 | 0003-6935 |
英文摘要 | after reviewing the spectrum-dividing principle of the savart polariscope (sp) in the polarization interference imaging spectrometer (piis) that we developed, we analyze the influences of the thickness, misalignment, and dispersion of the sp on the optical path difference (opd). the theoretical expression of the opd for the misalignment of the sp optical axis is deduced, and the opd is analyzed when the incident plane is parallel, at 45 degrees, or orthogonal to the principal section of the left plate of the sp. the selective thickness of the single savart plate is analyzed when it is placed at the ideal and misalignment positions. the influence of dispersion of the sp on the opd is analyzed when the misalignment error is +/- 1'. the relationships between the opd and wavelength are simulated and validated with experiments. this work can provide theoretical and practical guidance for the design, calibration, modulation, innovation, experiment, and engineering of the piis. (c) 2011 optical society of america |
WOS标题词 | science & technology ; physical sciences |
类目[WOS] | optics |
研究领域[WOS] | optics |
关键词[WOS] | upper atmospheric wind ; fourier-transform ; orientation |
收录类别 | SCI ; EI |
原文出处 | http://www.opticsinfobase.org/ao/abstract.cfm?uri=ao-50-20-3497 |
语种 | 英语 |
WOS记录号 | WOS:000293069000016 |
公开日期 | 2012-06-29 |
源URL | [http://ir.opt.ac.cn/handle/181661/19852] ![]() |
专题 | 西安光学精密机械研究所_检测技术研究中心 |
作者单位 | 1.Xi An Jiao Tong Univ, Sch Sci, Minist Educ, Key Lab Nonequilibrium Synth & Modulat Condensed, Xian 710049, Peoples R China 2.Chinese Acad Sci, Xian Inst Opt & Precis Mech, Xian 710119, Peoples R China 3.Mindray Med Int Ltd, Shenzhen 518057, Peoples R China |
推荐引用方式 GB/T 7714 | Zhang, Chunmin,Zhao, Jianke,Sun, Yao. Influences of the thickness, misalignment, and dispersion of the Savart polariscope on the optical path difference and spectral resolution in the polarization interference imaging spectrometer[J]. applied optics,2011,50(20):3497-3504. |
APA | Zhang, Chunmin,Zhao, Jianke,&Sun, Yao.(2011).Influences of the thickness, misalignment, and dispersion of the Savart polariscope on the optical path difference and spectral resolution in the polarization interference imaging spectrometer.applied optics,50(20),3497-3504. |
MLA | Zhang, Chunmin,et al."Influences of the thickness, misalignment, and dispersion of the Savart polariscope on the optical path difference and spectral resolution in the polarization interference imaging spectrometer".applied optics 50.20(2011):3497-3504. |