中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Performance test of steel tube enclosed thermistors in air temperature measurements (EI CONFERENCE)

文献类型:会议论文

作者Yang H.; Yan F.; Yan F.; Peng J.
出版日期2011
会议名称3rd International Conference on Measuring Technology and Mechatronics Automation, ICMTMA 2011, January 6, 2011 - January 7, 2011
会议地点Shanghai, China
关键词The performances of two steel tube enclosed precision thermistors are tested in a well controlled air environment to determine if the thermistors have enough accuracy for 0.01C levels of air temperature gradient measurement. Three Fluke's 5641 and three 5610 thermistors were tested for 24 hours in Fluke's 7008 calibration bath in the first step. In the second step the performances of these two thermistors in steady state and transient state air temperature environments were tested for 8 hours separately. The results illustrate that two of the 5641s were airproofed in a temperature well controlled tube when steel tube enclosed precision thermistors are used in air temperature measurement there are unstable departures from each other with a maximum 8.8 mC in constant air medium and 27 mC in temperature changing air medium. The transient-time errors are too large to show the true amounts of temperature changes for air temperature measurement. 2011 IEEE.
页码854-857
收录类别EI
源URL[http://ir.ciomp.ac.cn/handle/181722/33556]  
专题长春光学精密机械与物理研究所_中科院长春光机所知识产出_会议论文
推荐引用方式
GB/T 7714
Yang H.,Yan F.,Yan F.,et al. Performance test of steel tube enclosed thermistors in air temperature measurements (EI CONFERENCE)[C]. 见:3rd International Conference on Measuring Technology and Mechatronics Automation, ICMTMA 2011, January 6, 2011 - January 7, 2011. Shanghai, China.

入库方式: OAI收割

来源:长春光学精密机械与物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。