A new software method for accurate mass measurements of drugs on unit mass resolution mass spectrometer
文献类型:期刊论文
作者 | LIU Ke3; MA Bin3; WANG Yongdong2; CHEN Xiaoyan3![]() ![]() |
刊名 | Acta Pharmaceutica Sinica
![]() |
出版日期 | 2007 |
卷号 | 42期号:10页码:1112-1114 |
关键词 | accurate relative molecular mass isotope unit mass resolution mass spectrometer |
ISSN号 | 0513-4870 |
其他题名 | 一种新软件方法用于单位分辨质谱仪上药物相对分子质量的准确测定 |
文献子类 | Article |
英文摘要 | Accuratemassmeasurements of20 drugswere conducted usingMassWorks software on a TSQ Quantum Ultramass spectrometer. AllQ1 scanmass spectraldatawere collected in profilemodewith fullwidth athalf-maximum (FWHM) set at0·7 Da. The mass errors of all 20 drugs (molecularweight between 135 and 810) werewithin the range from -22·4*10-6to36·1*10-6, among them 90% of the drugs achieved a mass accuracy better than 10*10-6. The new method can provide accurate mass measurements on unitmass resolutionmass spectrometers. |
WOS研究方向 | Pharmacology & Pharmacy (provided by Clarivate Analytics) |
语种 | 中文 |
CSCD记录号 | CSCD:2927322 |
源URL | [http://119.78.100.183/handle/2S10ELR8/268166] ![]() |
专题 | 上海药物代谢研究中心 |
通讯作者 | ZHONG Dafang |
作者单位 | 1.Shanghai 201203, China. 2.Cerno Bioscience, Danbury, CT 06810, USA.; 3.Shanghai Institute of Materia Medica; |
推荐引用方式 GB/T 7714 | LIU Ke,MA Bin,WANG Yongdong,et al. A new software method for accurate mass measurements of drugs on unit mass resolution mass spectrometer[J]. Acta Pharmaceutica Sinica,2007,42(10):1112-1114. |
APA | LIU Ke,MA Bin,WANG Yongdong,CHEN Xiaoyan,&ZHONG Dafang.(2007).A new software method for accurate mass measurements of drugs on unit mass resolution mass spectrometer.Acta Pharmaceutica Sinica,42(10),1112-1114. |
MLA | LIU Ke,et al."A new software method for accurate mass measurements of drugs on unit mass resolution mass spectrometer".Acta Pharmaceutica Sinica 42.10(2007):1112-1114. |
入库方式: OAI收割