中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Development of Mechanostimulated Patch-Clamp System for Cellular Physiological Study

文献类型:期刊论文

作者Zhang, Changlin2,3; Li, Peng2; Liu, Lianqing2; Wang, Yuechao2; Gao, Zhaobing1; Li, Guangyong4
刊名IEEE-ASME TRANSACTIONS ON MECHATRONICS
出版日期2014-08
卷号19期号:4页码:1138-1147
关键词Atomic force microscope (AFM) mechanosensitive (MS) ion channel planar patch clamp
ISSN号1083-4435
DOI10.1109/TMECH.2013.2272562
文献子类Article
英文摘要Mechanosensitive ion channels play important roles for sensing and responding to the mechanical stimuli signals in living life. Here we report the development of a mechanostimulated patch-clamp system for simultaneous recording of external stimuli and acquisition of cellular physiological responses. This system integrates a custom-designed planar patch-clamp system with a robot-assisted atomic force microscope (AFM) system. The former, with a microfluidic channel, can realize not only recording electrical signals but also exchanging intracellular solution; while the latter, enhanced by robotic techniques (local scan force feedback, augmented reality vision feedback), can generate force stimuli with controllable patterns and magnitudes under the operator's real-time monitoring. To verify the performance of the developed system, we first measured the whole-cell current of the voltage-gated potassium ion channel Kv1.1 expressed on Human Embryonic Kidney (HEK293) cells and then recorded the mechanosensitive ion channel current in a mouse neuroblastoma cell line (Neuro2 A) in the whole-cell configuration during the AFM indenting on the membrane surface; finally, confirmed the ability to exchange intracellular solution by delivering propidium iodide into the captured cell through intracellular solution. The results prove the effectiveness of the system.
WOS关键词ATOMIC-FORCE MICROSCOPY ; GATED K+ CHANNELS ; MEMBRANE PATCHES ; SENSORY NEURONS ; SILICON CHIP ; ION CHANNELS ; LIVING CELLS ; WHOLE-CELL ; MECHANOTRANSDUCTION ; NANOMANIPULATION
资助项目National Natural Science Foundation of China[61175803] ; Instrument Developing Project of the Chinese Academy of Sciences[YZ201245] ; CAS/SAFEA International Partnership Program for Creative Research Teams[00000000] ; National Science Foundation[CNS 1035563]
WOS研究方向Automation & Control Systems ; Engineering
语种英语
WOS记录号WOS:000335915800004
出版者IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
源URL[http://119.78.100.183/handle/2S10ELR8/276976]  
专题神经药理学研究国际科学家工作站
通讯作者Zhang, Changlin
作者单位1.Chinese Acad Sci, Shanghai Inst Mat Med, Shanghai 201203, Peoples R China;
2.Chinese Acad Sci, State Key Lab Robot, Shenyang Inst Automat, Shenyang 110016, Peoples R China;
3.Univ Chinese Acad Sci, Beijing 100049, Peoples R China;
4.Univ Pittsburgh, Dept Elect & Comp Engn, Pittsburgh, PA 15213 USA
推荐引用方式
GB/T 7714
Zhang, Changlin,Li, Peng,Liu, Lianqing,et al. Development of Mechanostimulated Patch-Clamp System for Cellular Physiological Study[J]. IEEE-ASME TRANSACTIONS ON MECHATRONICS,2014,19(4):1138-1147.
APA Zhang, Changlin,Li, Peng,Liu, Lianqing,Wang, Yuechao,Gao, Zhaobing,&Li, Guangyong.(2014).Development of Mechanostimulated Patch-Clamp System for Cellular Physiological Study.IEEE-ASME TRANSACTIONS ON MECHATRONICS,19(4),1138-1147.
MLA Zhang, Changlin,et al."Development of Mechanostimulated Patch-Clamp System for Cellular Physiological Study".IEEE-ASME TRANSACTIONS ON MECHATRONICS 19.4(2014):1138-1147.

入库方式: OAI收割

来源:上海药物研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。