A high precision square law detector in digital microwave radiometer
文献类型:会议论文
| 作者 | Pei, Xin1,2 ; Chen, Maozheng1 ; Yan, Hao1 ; Li, Jian1 ; Liu, Xiaohong1,2
|
| 出版日期 | 2011-09-08 |
| 会议日期 | May 27, 2011 - May 29, 2011 |
| 会议地点 | Xi'an, China |
| 关键词 | Microwaves Communication Detectors Microwave devices Microwave measurement Radio astronomy Radiometers Radiometry Schottky barrier diodes |
| DOI | 10.1109/ICCSN.2011.6014425 |
| 页码 | 211-214 |
| 国家 | China |
| 英文摘要 | A new square law detector has been developed for digital microwave radiometer. It has a wider dynamic range and a more accurate square law response than has been available in the past. Experimental results show that the voltage sensitivity is higher than 0.9mV/uW and the linearity error is less than 0.3% in the frequency range from 100MHz to 2GHz and power range from 50dBm to 20dBm. Another advantage of the detector is its flatness, by selected a best type of diodes. Detailed detection circuit and measurement system have been given in this paper. 2011 IEEE. |
| 产权排序 | 1 |
| 会议录 | 2011 IEEE 3rd International Conference on Communication Software and Networks, ICCSN 2011
![]() |
| 会议录出版者 | IEEE Computer Society |
| 语种 | 英语 |
| URL标识 | 查看原文 |
| ISBN号 | 9781612844855 |
| 源URL | [http://ir.xao.ac.cn/handle/45760611-7/3142] ![]() |
| 专题 | 微波接收机技术实验室 |
| 通讯作者 | Pei, Xin |
| 作者单位 | 1.XinJiang Astronomical Observatory, Chinese Academy of Sciences, Urumqi, China; 2.Graduate University of Chinese Academy of Sciences, BeiJing, China |
| 推荐引用方式 GB/T 7714 | Pei, Xin,Chen, Maozheng,Yan, Hao,et al. A high precision square law detector in digital microwave radiometer[C]. 见:. Xi'an, China. May 27, 2011 - May 29, 2011. |