Stereotactic dose perturbation from an aneurysm clip measured by Gafchromic®EBT film
文献类型:期刊论文
作者 | He, CF (He, Cheng-fa)[ 2,4 ]![]() |
刊名 | Australasian Physical and Engineering Sciences in Medicine
![]() |
出版日期 | 2008 |
卷号 | 31期号:1页码:18-23 |
ISSN号 | 0158-9938 |
DOI | 10.1007/BF03178449 |
英文摘要 | Aneurysm clips within stereotactic treatment volumes enhance spatial accuracy but perturb the dose distribution. The dose perturbations caused by a standard titanium alloy aneurysm clip (Ti6Al4V) have been measured with Gafchromic® EBT film. The maximum dose perturbation was an increase of 6% within 0.5 mm of the beam entry surface of the clip, and a decrease of 7% within 0.5 mm of the beam exit surface of the clip. Results also showed perturbations to film readout due to the presence of micro dust particles on the film affecting optical properties at high spatial resolution (21um) scanning. Special procedures should be used when film is immersed in water, dried and then read at high spatial resolution. We recommend that films should be immersed only in distilled water and tools such as canned air puffs should be used to clean films without scratches. |
WOS记录号 | WOS:000260409500006 |
公开日期 | 2012-11-29 |
源URL | [http://ir.xjipc.cas.cn/handle/365002/1989] ![]() |
专题 | 新疆理化技术研究所_材料物理与化学研究室 固体辐射物理研究室 |
作者单位 | 1.Chinese Acad Sci, Xinjiang Tech Inst Phys & Chem, Urumqi, Xinjiang, Peoples R China 2.Alfred Hosp, William Buckland Radiotherapy Ctr, Melbourne, Vic, Australia 3.China Inst Atom Energy, Beijing, Peoples R China 4.RMIT Univ, Sch Med Sci, Bundoora, Vic 3083, Australia |
推荐引用方式 GB/T 7714 | He, CF ,Geso, M ,Ackerly, T ,et al. Stereotactic dose perturbation from an aneurysm clip measured by Gafchromic®EBT film[J]. Australasian Physical and Engineering Sciences in Medicine,2008,31(1):18-23. |
APA | He, CF ,Geso, M ,Ackerly, T ,&Wong, CJ .(2008).Stereotactic dose perturbation from an aneurysm clip measured by Gafchromic®EBT film.Australasian Physical and Engineering Sciences in Medicine,31(1),18-23. |
MLA | He, CF ,et al."Stereotactic dose perturbation from an aneurysm clip measured by Gafchromic®EBT film".Australasian Physical and Engineering Sciences in Medicine 31.1(2008):18-23. |
入库方式: OAI收割
来源:新疆理化技术研究所
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。