A portable data-collection system for soft x-ray absorption spectroscopy in the Shanghai Synchrotron Radiation Facility
文献类型:期刊论文
作者 | Sun, TX; Meng, XY; Cao, JF; Wang, Y; Guo, Z; Wang, ZJ; Liu, HG; Zhang, XZ; Tai, RZ |
刊名 | REVIEW OF SCIENTIFIC INSTRUMENTS
![]() |
出版日期 | 2020 |
卷号 | 91期号:1页码:- |
关键词 | XAFS TEMPERATURE |
ISSN号 | 0034-6748 |
DOI | 10.1063/1.5128054 |
文献子类 | 期刊论文 |
英文摘要 | Based on the Experimental Physics and Industrial Control System, a portable data-collection system for soft x-ray absorption spectroscopy has been developed at the BL02B and BL08U beamlines of the Shanghai Synchrotron Radiation Facility. The data-collection system can be used to carry out total electron yield (TEY) and total fluorescence yield (TFY) experiments simultaneously. The hardware consists of current preamplifiers, voltage-to-frequency converters, and a multi-channel counter, which are aimed at improving the signal-to-noise ratio. The control logic is developed using Python and Java. The novelty of this control system is its designed portability while being extensible and readable and having low noise and high real-time capabilities. The oxygen K-edge absorption spectra of SrTiO3 were obtained using the TEY and TFY technology at the BL02B beamline. Furthermore, the TEY and TFY spectra of the relaxor ferroelectric single-crystal of lead magnesium niobate-lead titanate measured by the present data-collection system have lower peak-to-peak noise amplitude than the ones measured by using a picoammeter. The experimental results show that the spectral signal-to-noise ratio recorded by the present system is 5.7-12.4 dB higher than that with the picoammeter detector. |
语种 | 英语 |
源URL | [http://ir.sinap.ac.cn/handle/331007/32686] ![]() |
专题 | 上海应用物理研究所_中科院上海应用物理研究所2011-2017年 |
作者单位 | 1.Chinese Acad Sci, Shanghai Adv Res Inst, Shanghai Synchrotron Radiat Facil, Shanghai 201210, Peoples R China 2.Chinese Acad Sci, Shanghai Inst Appl Phys, Shanghai 201800, Peoples R China 3.Univ Chinese Acad Sci, Beijing 200049, Peoples R China |
推荐引用方式 GB/T 7714 | Sun, TX,Meng, XY,Cao, JF,et al. A portable data-collection system for soft x-ray absorption spectroscopy in the Shanghai Synchrotron Radiation Facility[J]. REVIEW OF SCIENTIFIC INSTRUMENTS,2020,91(1):-. |
APA | Sun, TX.,Meng, XY.,Cao, JF.,Wang, Y.,Guo, Z.,...&Tai, RZ.(2020).A portable data-collection system for soft x-ray absorption spectroscopy in the Shanghai Synchrotron Radiation Facility.REVIEW OF SCIENTIFIC INSTRUMENTS,91(1),-. |
MLA | Sun, TX,et al."A portable data-collection system for soft x-ray absorption spectroscopy in the Shanghai Synchrotron Radiation Facility".REVIEW OF SCIENTIFIC INSTRUMENTS 91.1(2020):-. |
入库方式: OAI收割
来源:上海应用物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。