中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM

文献类型:预印本

作者Zhang, Dali; Li, Xinqiao; Wen, Xiangyang; Xiong, Shaolin; An, Zhenghua; Sun, Xilei; Qiao, Rui; Li, Zhengwei; Gong, Ke; Hou, Dongjie
学科主题Physics
关键词Instrumentation and Detectors High Energy Physics - Experiment
英文摘要Each satellite of the Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM, mission) consists of 25 SiPM based gamma-ray detectors (GRDs). Although SiPM based GRD has merits of compact size and low bias-voltage, the drift of the SiPM gain with temperature is a severe problem for GRD performance. An adaptive voltage supply source was designed to automatically adjust the SiPM bias voltage to compensate the temperature effects and keep the gain stable. This approach has been proved to be effective during both the on-ground and in-flight tests. The in-flight measured variation of the SiPM gain is within 2%. To reduce the gain non-uniformity of GRDs, an iterative bias voltage adjustment approach is proposed and implemented. The gain non-uniformity is reduced from 17% to 0.6%. In this paper, the gain stabilization and consistency correction approach are presented and discussed in detail.
发表日期2021
语种英语
源URL[http://ir.ihep.ac.cn/handle/311005/289906]  
专题引力波暴高能电磁对应体全天监测器(GECAM)
高能物理研究所_实验物理中心
高能物理研究所_粒子天体物理中心
高能物理研究所_多学科研究中心
高能物理研究所_管理与技术支持
高能物理研究所_东莞分部
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhang, Dali,Li, Xinqiao,Wen, Xiangyang,et al. Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM. 2021.

入库方式: OAI收割

来源:高能物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。