Geometric calibration method based on a two-dimensional turntable for a directional polarimetric camera
文献类型:期刊论文
作者 | Huang Chan2,3,4; Meng Binghuan2,4![]() ![]() ![]() |
刊名 | Applied Optics
![]() |
出版日期 | 2020-01 |
关键词 | DPC Geometric calibration two-dimensional turntable |
DOI | https://doi.org/10.1364/AO.59.000226 |
英文摘要 | The directional polarimetric camera (DPC) is a polarization sensor with ultra-wide-angle and low-distortion imaging characteristics. Geometric calibration is usually the first essential step before remote sensing satellites are launched. In this paper, a geometric calibration method based on a two-dimensional turntable and a rotation matrix with high precision, simple operation, and wide application range is proposed for the directional polari metric camera. Instead of precisely adjusting the position of the sensor on the turntable, the method effectively eliminates the errors caused by the mechanical axis of the turntable and the optical axis of the sensor not being adjusted to the same direction through the rotation transformation of the coordinate system. The geometric cali bration experiments of the directional polarimetric camera were carried out with the method of Chen et al. [Optik 121, 486 (2010)] and the proposed method. The experimental results showed the calibration residual of the pro posed method was less than 0.1 pixel while Chen’s method was 0.3 pixel. The mean reprojection error and root mean square error of the proposed method were reduced to 0.06352 pixel and 0.06961 pixel, respectively. The geo metric calibration parameters obtained by the proposed method were used for geometric correction of the in-orbit images of the DPC, and the results also prove the effectiveness and superiority of the proposed method. |
学科主题 | 航空、航天科学技术 |
语种 | 英语 |
源URL | [http://ir.hfcas.ac.cn:8080/handle/334002/132267] ![]() |
专题 | 合肥物质科学研究院_中科院安徽光学精密机械研究所 |
通讯作者 | Meng Binghuan |
作者单位 | 1.Shanghai Institute of Satellite Engineering, Shanghai 200240, China 2.3Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China 3.University of Science and Technology of China, Hefei, Anhui 230026, China 4.Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China, |
推荐引用方式 GB/T 7714 | Huang Chan,Meng Binghuan,Chang Yuyang,et al. Geometric calibration method based on a two-dimensional turntable for a directional polarimetric camera[J]. Applied Optics,2020. |
APA | Huang Chan.,Meng Binghuan.,Chang Yuyang.,Chen Feinan.,Zhang Miaomiao.,...&Hong Jin.(2020).Geometric calibration method based on a two-dimensional turntable for a directional polarimetric camera.Applied Optics. |
MLA | Huang Chan,et al."Geometric calibration method based on a two-dimensional turntable for a directional polarimetric camera".Applied Optics (2020). |
入库方式: OAI收割
来源:合肥物质科学研究院
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。