中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
On-line measurement of propofol using membrane inlet ion mobility spectrometer

文献类型:期刊论文

作者Zhou, Qinghua1,2; Wang, Weiguo1; Cang, Huaiwen1; Du, Yongzhai1,2; Han, Fenglei1,2; Chen, Chuang1,2; Cheng, Shasha1,2; Li, Jinghua1; Li, Haiyang1
刊名talanta
出版日期2012-08-30
卷号98页码:241-246
关键词Membrane inlet Ion mobility spectrometer Propofol Exhaled air
ISSN号0039-9140
产权排序1,1
通讯作者李海洋
英文摘要the concentration of propofol in patient's exhaled air is an indicator of the anesthetic depth. in the present study, a membrane inlet ion mobility spectrometer (mi-ims) was built for the on-line measurement of propofol. compared with the direct sample introduction, the membrane inlet could eliminate the interference of moisture and improve the selectivity of propofol. effects of membrane temperature and carrier gas flow rate on the sensitivity and response time have been investigated experimentally and theoretically. under the optimized experimental conditions of membrane temperature 100 degrees c and carrier gas flow rate 200 ml min(-1), the calculated limit of detection (lod) for propofol was 1 ppbv, and the calibration curve was linear in the range of 10-83 ppbv with a correlation coefficient (r-2) of 0.993. finally, the propofol concentration in an anaesthetized mouse exhaled air was monitored continuously to demonstrate the capability of mi-ims in the on-line measurement of propofol in real samples. (c) 2012 elsevier b.v. all rights reserved.
学科主题物理化学
类目[WOS]chemistry, analytical
研究领域[WOS]chemistry
关键词[WOS]degradation-products ; gas ; water ; feasibility ; metabolites ; explosives ; arterial ; surgery ; plasma ; air
收录类别SCI
语种英语
WOS记录号WOS:000309327600035
公开日期2013-10-11
源URL[http://159.226.238.44/handle/321008/118156]  
专题大连化学物理研究所_中国科学院大连化学物理研究所
作者单位1.Chinese Acad Sci, Dalian Inst Chem Phys, Dalian 116023, Liaoning, Peoples R China
2.Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
推荐引用方式
GB/T 7714
Zhou, Qinghua,Wang, Weiguo,Cang, Huaiwen,et al. On-line measurement of propofol using membrane inlet ion mobility spectrometer[J]. talanta,2012,98:241-246.
APA Zhou, Qinghua.,Wang, Weiguo.,Cang, Huaiwen.,Du, Yongzhai.,Han, Fenglei.,...&Li, Haiyang.(2012).On-line measurement of propofol using membrane inlet ion mobility spectrometer.talanta,98,241-246.
MLA Zhou, Qinghua,et al."On-line measurement of propofol using membrane inlet ion mobility spectrometer".talanta 98(2012):241-246.

入库方式: OAI收割

来源:大连化学物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。