Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging
文献类型:期刊论文
作者 | Li, Da2; Ren, Yong3![]() ![]() |
刊名 | IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
![]() |
出版日期 | 2024-10-01 |
卷号 | 34 |
关键词 | Coils Magnetic resonance imaging Superconducting magnets Magnetic fields Windings Stress Electromagnetic forces Head scanner magnetic resonance imaging (MRI) REBCO coil superconducting magnet |
ISSN号 | 1051-8223 |
DOI | 10.1109/TASC.2024.3426543 |
通讯作者 | Ren, Yong(renyong@mail.ustc.edu.cn) |
英文摘要 | A 0.5 T REBCO MRI coil with a warm bore of 560 mm was designed for head imaging using linear programming (LP) and nonlinear programming (NLP). The LP method was used to determine the initial distribution of the coil cluster, followed by the NLP method, which optimized the radii and heights of the coil blocks using this initial layout as a reference. Both layer-winding and double-pancake (DP) winding coils were explored for developing MRI magnets. The layer-winding coil enabled the MRI magnet to achieve a uniformity of 10 parts per million (ppm) over a 200-mm diameter spherical volume. In contrast, the DP winding coils attained a field uniformity of 17.6 ppm without employing either iron or superconducting shim coils. The stray magnetic field of the MRI magnet was designed to be within 0.5 mT at radial and axial positions of 3.2 and 4.2 m, respectively. Because the REBCO magnet has a substantial temperature margin, it can be effectively cooled using solid nitrogen at 35 K or through conduction cooling. This article presents the electromagnetic design and mechanical analysis results of the coils. |
资助项目 | Anhui Provincial Natural Science Foundation for Distinguished Young Scholars[2108085J27] ; China Scholarship Council[202004910193] |
WOS研究方向 | Engineering ; Physics |
语种 | 英语 |
WOS记录号 | WOS:001276387000003 |
出版者 | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC |
资助机构 | Anhui Provincial Natural Science Foundation for Distinguished Young Scholars ; China Scholarship Council |
源URL | [http://ir.hfcas.ac.cn:8080/handle/334002/137338] ![]() |
专题 | 中国科学院合肥物质科学研究院 |
通讯作者 | Ren, Yong |
作者单位 | 1.Univ Queensland, Sch Elect Engn & Comp Sci, Brisbane, Qld 4072, Australia 2.Univ Sci & Technol China, Hefei 230031, Peoples R China 3.Chinese Acad Sci, Hefei Inst Phys Sci, Inst Plasma Phys, Hefei 200031, Peoples R China |
推荐引用方式 GB/T 7714 | Li, Da,Ren, Yong,Liu, Feng. Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging[J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY,2024,34. |
APA | Li, Da,Ren, Yong,&Liu, Feng.(2024).Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging.IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY,34. |
MLA | Li, Da,et al."Electromagnetic Design of a 0.5 T REBCO MRI Magnet for Brain MRI Imaging".IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY 34(2024). |