Measurement of spatial resolution of the micro-CT system

文献类型:期刊论文

作者Rong, JY; Fu, GT; Wei, CF; Sun, CL; Wei, L
刊名CHINESE PHYSICS C
出版日期2010
卷号34期号:3页码:412-416
关键词micro-CT line-pair resolution MTF
ISSN号1674-1137
DOI10.1088/1674-1137/34/3/022
文献子类期刊
英文摘要The performance test is an important and necessary work for the micro-CT (computed tomography) system. The focal spot size of the micro focus X-ray tube is measured. The method of measuring the spatial resolution of micro-CT is introduced. A line-pair resolution of 28.2 lp/mm at the 10% modulation transfer function (MTF) level can be achieved with 14.7 mu m spot, size, 12.3 mu m voxel size and a 25 mm field of view. In addition, a tungsten wire with the diameter of 5 mu m can be detected by the system.
电子版国际标准刊号2058-6132
WOS研究方向Physics
语种英语
WOS记录号WOS:000275495100022