Evaluation of imaging protocol for ECT based on CS image reconstruction algorithm

文献类型:期刊论文

作者Zhou, XL; Yun, MK; Cao, XX; Liu, SQ; Wang, L; Huang, XC; Wei, L
刊名CHINESE PHYSICS C
出版日期2014
卷号38期号:4页码:048201
关键词ECT imaging protocol compressed sensing total variation
ISSN号1674-1137
DOI10.1088/1674-1137/38/4/048201
文献子类Article
英文摘要

Single-photon emission computerized tomography and positron emission tomography are essential medical imaging tools, for which the sampling angle number and scan time should be carefully chosen to give a good compromise between image quality and radiopharmaceutical dose. In this study, the image quality of different acquisition protocols was evaluated via varied angle number and count number per angle with Monte Carlo simulation data. It was shown that, when similar imaging counts were used, the factor of acquisition counts was more important than that of the sampling number in emission computerized tomography. To further reduce the activity requirement and the scan duration, an iterative image reconstruction algorithm for limited-view and low-dose tomography based on compressed sensing theory has been developed. The total variation regulation was added to the reconstruction process to improve the signal to noise Ratio and reduce artifacts caused by the limited angle sampling. Maximization of the maximum likelihood of the estimated image and the measured data and minimization of the total variation of the image are alternatively implemented. By using this advanced algorithm, the reconstruction process is able to achieve image quality matching or exceed that of normal scans with only half of the injection radiopharmaceutical dose.

电子版国际标准刊号2058-6132
WOS关键词SCAN DURATION ; PROJECTION ; PET/CT ; IMPACT ; SPECT
WOS研究方向Physics
语种英语
WOS记录号WOS:000333547500022
源URL[https://ir.ihep.ac.cn/handle/311005/304255]  
专题高能物理研究所_核技术应用研究中心
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhou, XL,Yun, MK,Cao, XX,et al. Evaluation of imaging protocol for ECT based on CS image reconstruction algorithm[J]. CHINESE PHYSICS C,2014,38(4):048201.
APA Zhou, XL.,Yun, MK.,Cao, XX.,Liu, SQ.,Wang, L.,...&Wei, L.(2014).Evaluation of imaging protocol for ECT based on CS image reconstruction algorithm.CHINESE PHYSICS C,38(4),048201.
MLA Zhou, XL,et al."Evaluation of imaging protocol for ECT based on CS image reconstruction algorithm".CHINESE PHYSICS C 38.4(2014):048201.

入库方式: OAI收割

来源:高能物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。