7T Transmit/Receive Arrays Using ICE Decoupling for Human Head MR Imaging
文献类型:期刊论文
| 作者 | Yan, XQ; Zhang, XL; Feng, BT; Ma, CX; Wei, L; Xue, R |
| 刊名 | IEEE TRANSACTIONS ON MEDICAL IMAGING
![]() |
| 出版日期 | 2014 |
| 卷号 | 33期号:9页码:1781-1787 |
| 关键词 | Decoupling high field magnetic resonance imaging (MRI) human head induced current compensation or elimination (ICE) loop parallel imaging phased array radio-frequency (RF) coil |
| ISSN号 | 0278-0062 |
| DOI | 10.1109/TMI.2014.2313879 |
| 文献子类 | Article |
| 英文摘要 | In designing large-sized volume type phased array coils for human head imaging at ultrahigh fields, e.g., 7T, minimizing electromagnetic coupling among array elements is technically challenging. A new decoupling method based on induced current compensation or elimination (ICE) for a microstrip line planar array has recently been proposed. In this study, an eight-channel transmit/receive volume array with ICE-decoupled loop elements was built and investigated to demonstrate its feasibility and robustness for human head imaging at 7T. Isolation between adjacent loop elements was better than - 25 dB with a human head load. The worst-case of the isolation between all of the elements was about - 17.5 dB. All of the MRI experiments were performed on a 7T whole-body human MR scanner. Images of the phantom and human head were acquired and g-factor maps were measured and calculated to evaluate the performance of the coil array. Compared with the conventional capacitively decoupled array, the ICE-decoupled array demonstrated improved parallel imaging ability and had a higher SNR. The experimental results indicate that the transceiver array design with ICE decoupling technique might be a promising solution to designing high performance transmit/receive coil arrays for human head imaging at ultrahigh fields. |
| 电子版国际标准刊号 | 1558-254X |
| WOS关键词 | PARALLEL MRI ; COIL ARRAY ; PHASED-ARRAY ; HUMAN BRAIN ; RF POWER ; DESIGN ; FIELD ; HARMONICS ; SMASH ; TESLA |
| WOS研究方向 | Computer Science ; Engineering ; Imaging Science & Photographic Technology ; Radiology, Nuclear Medicine & Medical Imaging |
| 语种 | 英语 |
| WOS记录号 | WOS:000341773600002 |
| 源URL | [https://ir.ihep.ac.cn/handle/311005/304263] ![]() |
| 专题 | 高能物理研究所_核技术应用研究中心 |
| 作者单位 | 中国科学院高能物理研究所 |
| 推荐引用方式 GB/T 7714 | Yan, XQ,Zhang, XL,Feng, BT,et al. 7T Transmit/Receive Arrays Using ICE Decoupling for Human Head MR Imaging[J]. IEEE TRANSACTIONS ON MEDICAL IMAGING,2014,33(9):1781-1787. |
| APA | Yan, XQ,Zhang, XL,Feng, BT,Ma, CX,Wei, L,&Xue, R.(2014).7T Transmit/Receive Arrays Using ICE Decoupling for Human Head MR Imaging.IEEE TRANSACTIONS ON MEDICAL IMAGING,33(9),1781-1787. |
| MLA | Yan, XQ,et al."7T Transmit/Receive Arrays Using ICE Decoupling for Human Head MR Imaging".IEEE TRANSACTIONS ON MEDICAL IMAGING 33.9(2014):1781-1787. |
入库方式: OAI收割
来源:高能物理研究所
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。