A new method to determine the center of rotation shift in 2D-CT scanning system using image cross correlation

文献类型:期刊论文

作者Yang, Min1; Gao, Haidong1; Li, Xingdong2; Meng, Fanyong3; Wei, Dongbo1
刊名NDT & E INTERNATIONAL
出版日期2012-03-01
卷号46期号:0页码:48-54
关键词Computed tomography Center of rotation Image cross correlation Artifacts
ISSN号0963-8695
通讯作者Yang, M
英文摘要Center of rotation (COR) is one of the most important factors that affect the image quality of CT system. In this paper we propose a new method that is based on image cross correlation to accurately measure the shift of COR in 2D-CT (Two Dimensional Computed Tomography) system. In this method, we firstly rebin the fan-beam sinogram to parallel beam sinogram. Based on the symmetry characteristic of two projections that are 180 opposed to one another in the parallel beam sinogram, we then divide the rebinned sinogram equally into two sub-sinograms along rotating angle direction. After that we flip the second sub-sinogram in right direction, and apply the cross correlation operation to the first sub-sinogram and the flipped sub-sinogram to get the shift of COR. This method does not need the calibration sample, and just uses the sinogram of the scanned slice to calculate the shift of COR directly. The experimental results prove that this method is of high accuracy and anti-noise ability and is easy to implement. (C) 2011 Published by Elsevier Ltd.
WOS标题词Science & Technology ; Technology
类目[WOS]Materials Science, Characterization & Testing
研究领域[WOS]Materials Science
关键词[WOS]COMPUTED-TOMOGRAPHY
收录类别SCI
语种英语
WOS记录号WOS:000300547800007
公开日期2013-10-26
版本出版稿
源URL[http://ir.ipe.ac.cn/handle/122111/4292]  
专题过程工程研究所_多相复杂系统国家重点实验室
作者单位1.Beijing Univ Aeronaut & Astronaut, Sch Mech Engn & Automat, Beijing 100191, Peoples R China
2.Natl Inst Metrol, Div Metrol Ionizing Radiat & Med, Beijing 100013, Peoples R China
3.Chinese Acad Sci, Inst Proc Engn, State Key Lab Multiphase Complex Syst, Beijing 100190, Peoples R China
推荐引用方式
GB/T 7714
Yang, Min,Gao, Haidong,Li, Xingdong,et al. A new method to determine the center of rotation shift in 2D-CT scanning system using image cross correlation[J]. NDT & E INTERNATIONAL,2012,46(0):48-54.
APA Yang, Min,Gao, Haidong,Li, Xingdong,Meng, Fanyong,&Wei, Dongbo.(2012).A new method to determine the center of rotation shift in 2D-CT scanning system using image cross correlation.NDT & E INTERNATIONAL,46(0),48-54.
MLA Yang, Min,et al."A new method to determine the center of rotation shift in 2D-CT scanning system using image cross correlation".NDT & E INTERNATIONAL 46.0(2012):48-54.

入库方式: OAI收割

来源:过程工程研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。