Mechanism of peak drift of a grating monochromator and a designed sequential inductively coupled plasma spectrometer with an intelligent wavelength calibrating device

文献类型:期刊论文

作者Liu, K. L. ; Tan, H. B. ; Huang, Z. ; Huang, M.
刊名Journal of Analytical Atomic Spectrometry
出版日期1996
期号11页码:1019-1022
关键词monochromator sequential inductively coupled plasma spectrometer wavelength calibration
其他题名J. Anal. At. Spectrom.
中文摘要The mechanism of the peak drift of exit spectral lines of a grating monochromator was studied experimentally and theoretically, The results indicate that the peak drift is due not only to temperature change, but also to wavelength change and systematic errors, Equations describing the non-linear functional relationship between peak drift, wavelength and temperature were obtained. Based on these equations, a computer wavelength-calibrating program was developed, and an intelligent wavelength-calibrating device was designed for use with a sequential ICP spectrometer instead of the conventional local thermostat or the real-time line-profile method. The sequential ICP spectrometer with the intelligent wavelength-calibrating device (designated the ICP-IWC) was used to analyse various standards. The analytical results mere in good agreement with the certified values, and the RSDs of the determinations were about 1%. Compared with the use of the conventional local thermostat or the line-profile method, the ICP-IWC has been shown to be effective; the IWC device is easy to construct and should find wide application.
收录类别SCI
原文出处://WOS:A1996VT96500003
语种英语
公开日期2013-11-15
版本出版稿
源URL[http://ir.ipe.ac.cn/handle/122111/5990]  
专题过程工程研究所_研究所(批量导入)
推荐引用方式
GB/T 7714
Liu, K. L.,Tan, H. B.,Huang, Z.,et al. Mechanism of peak drift of a grating monochromator and a designed sequential inductively coupled plasma spectrometer with an intelligent wavelength calibrating device[J]. Journal of Analytical Atomic Spectrometry,1996(11):1019-1022.
APA Liu, K. L.,Tan, H. B.,Huang, Z.,&Huang, M..(1996).Mechanism of peak drift of a grating monochromator and a designed sequential inductively coupled plasma spectrometer with an intelligent wavelength calibrating device.Journal of Analytical Atomic Spectrometry(11),1019-1022.
MLA Liu, K. L.,et al."Mechanism of peak drift of a grating monochromator and a designed sequential inductively coupled plasma spectrometer with an intelligent wavelength calibrating device".Journal of Analytical Atomic Spectrometry .11(1996):1019-1022.

入库方式: OAI收割

来源:过程工程研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。