Humidity-Induced Spectral Shift in a Cross-Dispersion Echelle Spectrometer and Its Theoretical Investigation

文献类型:期刊论文

作者Liu Ke-ling1; Huang Mao2; Gary, Hieftje M.3
刊名SPECTROSCOPY AND SPECTRAL ANALYSIS
出版日期2010-09-01
卷号30期号:9页码:2555-2559
关键词Emission spectrometry Position shift of a spectral line Spectral-shift equation
ISSN号1000-0593
其他题名Spectrosc. Spectr. Anal.
中文摘要The relationship between ambient relative humidity H and the position shift of a spectral line was investigated both experimentally and theoretically. An echelle-based ICP emission spectrometer equipped with a CID detector was used for experimental verification of the derived model. The shift of a spectral line is quantitatively described by two defined spectral shift functions: Delta lambda(x) (x, lambda, H) (in the x direction of the CID detector) and Delta lambda(y) (y, lambda, H) (in the y direction of the CID detector). Experimental results indicate that Delta lambda(x) (x, lambda, H) does not change with a variation in ambient relative humidity, but Delta lambda(y) (y, lambda, H) does. A spectral shift equation, i. e. an empirical second-order polynomial equation, can be used to describe the relationship between Delta lambda(y) (y, lambda, H) and H. Based on the classical dipole model, classical mechanics and electrodynamics the empirical spectral-shift equation involving Delta lambda(y) (y, lambda, H) and H was theoretically deduced. The theoretical result is in good agreement with the experimental findings. The theoretical result's indicate that the coefficients of the empirical spectral-shift equation are related to the basic physical parameters of materials and the geometric configuration of the echelle CID ICP-AES, and also provide physical meaning to the coefficients of the empirical shift equation obtained experimentally.
英文摘要The relationship between ambient relative humidity H and the position shift of a spectral line was investigated both experimentally and theoretically. An echelle-based ICP emission spectrometer equipped with a CID detector was used for experimental verification of the derived model. The shift of a spectral line is quantitatively described by two defined spectral shift functions: Delta lambda(x) (x, lambda, H) (in the x direction of the CID detector) and Delta lambda(y) (y, lambda, H) (in the y direction of the CID detector). Experimental results indicate that Delta lambda(x) (x, lambda, H) does not change with a variation in ambient relative humidity, but Delta lambda(y) (y, lambda, H) does. A spectral shift equation, i. e. an empirical second-order polynomial equation, can be used to describe the relationship between Delta lambda(y) (y, lambda, H) and H. Based on the classical dipole model, classical mechanics and electrodynamics the empirical spectral-shift equation involving Delta lambda(y) (y, lambda, H) and H was theoretically deduced. The theoretical result is in good agreement with the experimental findings. The theoretical result's indicate that the coefficients of the empirical spectral-shift equation are related to the basic physical parameters of materials and the geometric configuration of the echelle CID ICP-AES, and also provide physical meaning to the coefficients of the empirical shift equation obtained experimentally.
WOS标题词Science & Technology ; Technology
类目[WOS]Spectroscopy
研究领域[WOS]Spectroscopy
收录类别SCI
原文出处://WOS:000281661000057
语种英语
WOS记录号WOS:000281661000057
公开日期2013-11-28
版本出版稿
源URL[http://ir.ipe.ac.cn/handle/122111/6366]  
专题过程工程研究所_研究所(批量导入)
作者单位1.Chinese Acad Sci, Inst Proc Engn, Beijing 100190, Peoples R China
2.Lib Chinese Acad Sci, Beijing 100190, Peoples R China
3.Indiana Univ, Dept Chem, Bloomington, IN 47405 USA
推荐引用方式
GB/T 7714
Liu Ke-ling,Huang Mao,Gary, Hieftje M.. Humidity-Induced Spectral Shift in a Cross-Dispersion Echelle Spectrometer and Its Theoretical Investigation[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS,2010,30(9):2555-2559.
APA Liu Ke-ling,Huang Mao,&Gary, Hieftje M..(2010).Humidity-Induced Spectral Shift in a Cross-Dispersion Echelle Spectrometer and Its Theoretical Investigation.SPECTROSCOPY AND SPECTRAL ANALYSIS,30(9),2555-2559.
MLA Liu Ke-ling,et al."Humidity-Induced Spectral Shift in a Cross-Dispersion Echelle Spectrometer and Its Theoretical Investigation".SPECTROSCOPY AND SPECTRAL ANALYSIS 30.9(2010):2555-2559.

入库方式: OAI收割

来源:过程工程研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。