A practical method to calibrate the slant angle of central X-ray for laminography scanning system

文献类型:期刊论文

作者Yang, Min1; Zhu, Jianhua1; Liu, Qiu1; Duan, Shengling1; Liang, Lihong2; Li, Xingdong3; Liu, Wenli3; Meng, Fanyong4
刊名NDT & E INTERNATIONAL
出版日期2014-06-01
卷号64期号:1页码:13-20
关键词Computed laminography (CL) Geometrical calibration Slant angle Ellipse fitting
ISSN号0963-8695
其他题名NDT E Int.
中文摘要Computed laminography (CL) has a particular function in plate-type structure tomography. For CL reconstruction, slant angle of the central X-ray is an important parameter and must be calibrated accurately. A practical feasible calibration method to determine the slant angle is proposed. According to this method, a spherical phantom is fixed on the rotary table and turns a full 360 rotation. Then, image processing and the least-square fitting are performed to figure out the locus of the projection. Consequently, the slant angle is achieved in accordance with the property of the fitted ellipse. The experimental results prove that this method is easy to implement and meet the inspection requirements. (C) 2014 Elsevier Ltd. All rights reserved.
英文摘要Computed laminography (CL) has a particular function in plate-type structure tomography. For CL reconstruction, slant angle of the central X-ray is an important parameter and must be calibrated accurately. A practical feasible calibration method to determine the slant angle is proposed. According to this method, a spherical phantom is fixed on the rotary table and turns a full 360 rotation. Then, image processing and the least-square fitting are performed to figure out the locus of the projection. Consequently, the slant angle is achieved in accordance with the property of the fitted ellipse. The experimental results prove that this method is easy to implement and meet the inspection requirements. (C) 2014 Elsevier Ltd. All rights reserved.
WOS标题词Science & Technology ; Technology
类目[WOS]Materials Science, Characterization & Testing
研究领域[WOS]Materials Science
关键词[WOS]CONE-BEAM CT ; COMPUTED-TOMOGRAPHY ; RECONSTRUCTION ; PARAMETERS ; PROJECTION ; GEOMETRY
收录类别SCI
原文出处://WOS:000335873700003
语种英语
WOS记录号WOS:000335873700003
公开日期2014-08-28
版本出版稿
源URL[http://ir.ipe.ac.cn/handle/122111/11008]  
专题过程工程研究所_研究所(批量导入)
作者单位1.Beijing Univ Aeronaut & Astronaut, Sch Mech Engn & Automat, Beijing 100191, Peoples R China
2.China Special Equipment Inspect & Res Inst, Beijing 100013, Peoples R China
3.Natl Inst Metrol, Beijing 100013, Peoples R China
4.Chinese Acad Sci, Inst Proc Engn, State Key Lab Multiphase Complex Syst, Beijing 100190, Peoples R China
推荐引用方式
GB/T 7714
Yang, Min,Zhu, Jianhua,Liu, Qiu,et al. A practical method to calibrate the slant angle of central X-ray for laminography scanning system[J]. NDT & E INTERNATIONAL,2014,64(1):13-20.
APA Yang, Min.,Zhu, Jianhua.,Liu, Qiu.,Duan, Shengling.,Liang, Lihong.,...&Meng, Fanyong.(2014).A practical method to calibrate the slant angle of central X-ray for laminography scanning system.NDT & E INTERNATIONAL,64(1),13-20.
MLA Yang, Min,et al."A practical method to calibrate the slant angle of central X-ray for laminography scanning system".NDT & E INTERNATIONAL 64.1(2014):13-20.

入库方式: OAI收割

来源:过程工程研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。