An active beamstop for accurate measurement of high intensity X-ray beams
文献类型:期刊论文
作者 | Pan, QY ; Wang, QS ; Wang, ZJ ; Li, L ; He, JH |
刊名 | NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
![]() |
出版日期 | 2014 |
卷号 | 735页码:584—586 |
关键词 | PROTEIN CRYSTALS RADIATION-DAMAGE |
ISSN号 | 0168-9002 |
通讯作者 | panqiangyan@sinap.ac.cn |
英文摘要 | In order to accurately measure the X-ray flux and determine the dose absorbed by the sample, we have designed an active beamstop device. Based on the photoelectric effect, the device consists of a beamstop, an insulator, and a concentric photoelectron collector. Tests of this device on the macromolecular crystallography beamline (BL17U1) at the Shanghai Synchrotron Research Facility (SSRF) show that the active beamstop is sufficient for determining the X-ray flux with the relative error smaller than 0.2%. The device can be used as a non-invasive, real-time intensity monitor at the sample during data collection. It can also be used with standard collimated beams or slits and served as a warning when the beam alignment drifts. Published by Elsevier By. |
收录类别 | SCI |
语种 | 英语 |
WOS记录号 | WOS:000327494600078 |
公开日期 | 2015-03-13 |
源URL | [http://ir.sinap.ac.cn/handle/331007/14005] ![]() |
专题 | 上海应用物理研究所_中科院上海应用物理研究所2011-2017年 |
推荐引用方式 GB/T 7714 | Pan, QY,Wang, QS,Wang, ZJ,et al. An active beamstop for accurate measurement of high intensity X-ray beams[J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,2014,735:584—586. |
APA | Pan, QY,Wang, QS,Wang, ZJ,Li, L,&He, JH.(2014).An active beamstop for accurate measurement of high intensity X-ray beams.NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,735,584—586. |