Ultrasensitive determination of cadmium in seawater by hollow fiber supported liquid membrane extraction coupled with graphite furnace atomic absorption spectrometry

文献类型:期刊论文

作者Peng, Jin-feng ; Liu, Rui ; Liu, Jing-fu ; He, Bin ; Hu, Xia-lin ; Jiang, Gui-bin
刊名SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
出版日期2007-05
卷号62期号:5页码:499-503
关键词cadmium seawater hollow fiber supported liquid membrane extraction graphite furnace atomic absorption spectrometry
英文摘要A new procedure, based on hollow fiber supported liquid membrane preconcentration coupled with graphite furnace atomic absorption spectrometry (GFAAS) detection, was developed for the determination of trace Cd in seawater samples. With 1-octanol that contained a mixture of dithizone (carrier) and oleic acid immobilized in the pores of the polypropylene hollow fiber as a liquid membrane, Cd was selectively extracted from water samples into 0.05 M HNO3 that filled the lumen of the hollow fiber as a stripping solution. The main extraction related parameters were optimized, and the effects of salinity and some coexisting interferants were also evaluated. Under the optimum extraction conditions, an enrichment factor of 3 87 was obtained for a I 00-mL sample solution. In combination with graphite furnace atomic absorption spectrometry, a very low detection limit (0.8 ng L-1) and a relative standard deviation (2.5% at 50 ng L-1 level) were achieved. Five seawater samples were analyzed by the proposed method without dilution, with detected Cd concentration in the range of 56.4-264.8 ng L-1 and the relative spiked recoveries over 89%. For comparison, these samples were also analyzed by the Inductively Coupled Plasma Mass Spectrometry (ICP-MS) method after a 10-fold dilution for matrix effect elimination. Statistical analysis with a one-way ANOVA shows no significant differences (at 0.05 level) between the results obtained by the proposed and ICP-MS methods. Additionally, analysis of certified reference materials (GBW (E) 080040) shows good agreement with the certified value. These results indicate that this present method is very sensitive and reliable, and can effectively eliminate complex matrix interferences in seawater samples. (C) 2007 Elsevier B.V. All rights reserved.
WOS记录号WOS:000248463000011
源URL[http://ir.rcees.ac.cn/handle/311016/22515]  
专题生态环境研究中心_环境化学与生态毒理学国家重点实验室
推荐引用方式
GB/T 7714
Peng, Jin-feng,Liu, Rui,Liu, Jing-fu,et al. Ultrasensitive determination of cadmium in seawater by hollow fiber supported liquid membrane extraction coupled with graphite furnace atomic absorption spectrometry[J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY,2007,62(5):499-503.
APA Peng, Jin-feng,Liu, Rui,Liu, Jing-fu,He, Bin,Hu, Xia-lin,&Jiang, Gui-bin.(2007).Ultrasensitive determination of cadmium in seawater by hollow fiber supported liquid membrane extraction coupled with graphite furnace atomic absorption spectrometry.SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY,62(5),499-503.
MLA Peng, Jin-feng,et al."Ultrasensitive determination of cadmium in seawater by hollow fiber supported liquid membrane extraction coupled with graphite furnace atomic absorption spectrometry".SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY 62.5(2007):499-503.

入库方式: OAI收割

来源:生态环境研究中心

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。