Multi-parameters decoupling method with Lamb wave sensor for improving the selectivity of label-free liquid detection (EI CONFERENCE)
会议论文
OAI收割
2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, TRANSDUCERS'11, June 5, 2011 - June 9, 2011, Beijing, China
作者:
Zhou L.
收藏
  |  
浏览/下载:24/0
  |  
提交时间:2013/03/25
The anti-symmetric modes (A01 mode for low frequency
A 03 mode for high frequency) and symmetric modes (S0 mode) produced by Lamb wave sensor are used to detect multi-parameters of a liquid
such as its density
sound velocity and viscosity. These modes are found to be very different to the parameters. For example
the A01 mode is very sensitive to the liquid's density but the A03 mode is sensitive to its sound velocity. The measurements of the attenuation with S0 mode can give out liquid's viscosity after its density been determined by A 01 mode. That could be a way to distinguish an unknown liquid with high sensitivity or to solve the problem of the selectivity of label free detection on biosensors. 2011 IEEE.