Improving photoacoustic-imaging axial positioning accuracy and signal-to-noise ratio using acoustic echo effect
文献类型:期刊论文
作者 | Zhao WX(赵文秀)2,3,4![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
刊名 | Sensors and Actuators A: Physical
![]() |
出版日期 | 2021 |
卷号 | 329页码:1-6 |
关键词 | Photoacoustic Echo effect Distance accuracy Signal-to-noise ratioa |
ISSN号 | 0924-4247 |
产权排序 | 1 |
英文摘要 | We have developed a method to improve the performance of photoacoustic (PA) imaging technique by applying the acoustic echo effect. By superimposing the main and echo ultrasound signals, the axial positioning accuracy of a PA imaging system could be improved by 100 % and the signal-to-noise ratio (SNR) could be enhanced by ∼2.8 dB experimentally. Using a multiwall carbon nanotubes composite structure as a sample, the occurrence conditions and properties of the echo signals were discussed. Theoretically, the SNR could be increased within a working distance of ∼3.9 mm when water is used as the medium. Moreover, we found that PA images of red blood cells could also be enhanced by this method. |
语种 | 英语 |
WOS记录号 | WOS:000659484700004 |
资助机构 | National Natural Science Foundation of China (Project Nos.61727811, 61573341, 61803366, 91748212, and U1613220) ; External Cooperation Program of the Chinese Academy of Sciences(Project No. 173321KYSB20170015) ; Youth Innovation Promo-tion Association CAS (Project No. Y201943) ; Hong Kong Research Grants Council (Project Nos. 9042639 and 907002) |
源URL | [http://ir.sia.cn/handle/173321/28863] ![]() |
专题 | 沈阳自动化研究所_机器人学研究室 |
通讯作者 | Liu LQ(刘连庆); Li WJ(李文荣) |
作者单位 | 1.Department of Mechanical Engineering, City University of Hong Kong, Kowloon Tong, Hong Kong 2.State Key Laboratory of Robotics, Shenyang Institute of Automation, Chinese Academy of Sciences, Shenyang 110016, China 3.Institutes for Robotics and Intelligent Manufacturing, Chinese Academy of Science, Shenyang 110016, China 4.University of Chinese Academy of Sciences, Beijing 100049, China 5.College of Mechanical and Electrical Engineering, Hohai University, Jiangsu 213022, China 6.Department of Chemistry and Bio-X, Stanford University, Stanford, CA 94305, United States 7.CAS-CityU Joint Laboratory on Robotics, City University of Hong Kong, Kowloon Tong, Hong Kong |
推荐引用方式 GB/T 7714 | Zhao WX,Yu HB,Wen YD,et al. Improving photoacoustic-imaging axial positioning accuracy and signal-to-noise ratio using acoustic echo effect[J]. Sensors and Actuators A: Physical,2021,329:1-6. |
APA | Zhao WX.,Yu HB.,Wen YD.,Li P.,Wang XD.,...&Li WJ.(2021).Improving photoacoustic-imaging axial positioning accuracy and signal-to-noise ratio using acoustic echo effect.Sensors and Actuators A: Physical,329,1-6. |
MLA | Zhao WX,et al."Improving photoacoustic-imaging axial positioning accuracy and signal-to-noise ratio using acoustic echo effect".Sensors and Actuators A: Physical 329(2021):1-6. |
入库方式: OAI收割
来源:沈阳自动化研究所
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。