中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Simultaneous Quantification of Total Antibody and Conjugated Payload for DS001 in Rat Serum Using a Hybrid Immuno-Capture LC-MS/MS

文献类型:期刊论文

作者Yu, Xiong3,4; Li, Weiqiang2,3; Huang, Wensi2,3; Xiao, Bo3,4; Long, Jing3; Wang, Qi3; Wang, Guifeng3; Wang, Chunhe3; Yu, Mingming1,4; Yu, Jinghua3
刊名AAPS JOURNAL
出版日期2025-01-07
卷号27期号:1页码:13
关键词ADC LC-MS/MS payload simultaneous quantification total antibody
ISSN号1550-7416
DOI10.1208/s12248-024-01007-4
通讯作者Yu, Mingming(yumingming@ouc.edu.cn) ; Yu, Jinghua(yujinghua2@simm.ac.cn) ; Diao, Xingxing(xxdiao@simm.ac.cn)
英文摘要Antibody-drug conjugates (ADCs) are intricate compounds that pose significant challenges in bioanalytical characterization. Therefore, multiple bioanalytical methods are required to comprehensively elucidate their pharmacokinetic (PK) profiles. In this study, we investigated DS001, an ADC consisting of a humanized monoclonal antibody (hRS7), a cleavable chemical linker, and the microtubule inhibitor monomethyl auristatin E (MMAE), with a drug-to-antibody ratio (DAR) of 8. This study established a rapid and sensitive hybrid immunoaffinity liquid-chromatography-tandem-mass-spectrometry (LC-MS/MS) approach for the simultaneous quantification of the total antibody and the enzymatically cleavable conjugated payload of DS001. This method is capable of monitoring fluctuations in average DAR values during PK assessments. The sample preparation procedure involved immunocapture, denaturation, trypsin digestion, papain digestion, and termination, all completed within a total processing time of less than 4 h. The method demonstrated linearity for the total antibody in the range of 100 ng/mL (lower-limit-of-quantification, LLOQ) to 100,000 ng/mL, and for the conjugated payload from 3.495 ng/mL (LLOQ) to 3495 ng/mL in rat serum. Both analytes exhibited standard curve correlation coefficients (r) greater than 0.990 within their respective linear ranges. The precision and accuracy of the method were within +/- 15% (+/- 20% for LLOQ). The verified LC-MS/MS approach was successfully employed in the PK analysis following intravenous administration of 0.2 mg/kg DS001 in rats via tail vein injection.
WOS关键词BIOANALYTICAL ASSAYS ; DRUG ; PHARMACOKINETICS ; VALIDATION ; STRATEGIES
资助项目National Natural Science Foundation of China[82373938] ; National Natural Science Foundation of China[82104275] ; National Natural Science Foundation of China[82104276] ; National Natural Science Foundation of China[82204585] ; Key Technologies R&D Program of Guangdong Province[2023B1111030004] ; National Key R&D Program of China[2022YFF1202600]
WOS研究方向Pharmacology & Pharmacy
语种英语
WOS记录号WOS:001391660200002
出版者SPRINGER
源URL[http://119.78.100.183/handle/2S10ELR8/315612]  
专题中国科学院上海药物研究所
通讯作者Yu, Mingming; Yu, Jinghua; Diao, Xingxing
作者单位1.Qingdao Natl Lab Marine Sci & Technol, Lab Marine Drugs & Bioprod, Qingdao 266003, Peoples R China
2.Univ Chinese Acad Sci, Beijing 100049, Peoples R China
3.Chinese Acad Sci, Shanghai Inst Mat Med, Shanghai 201210, Peoples R China
4.Ocean Univ China, Sch Med & Pharm, Qingdao 266003, Peoples R China
推荐引用方式
GB/T 7714
Yu, Xiong,Li, Weiqiang,Huang, Wensi,et al. Simultaneous Quantification of Total Antibody and Conjugated Payload for DS001 in Rat Serum Using a Hybrid Immuno-Capture LC-MS/MS[J]. AAPS JOURNAL,2025,27(1):13.
APA Yu, Xiong.,Li, Weiqiang.,Huang, Wensi.,Xiao, Bo.,Long, Jing.,...&Diao, Xingxing.(2025).Simultaneous Quantification of Total Antibody and Conjugated Payload for DS001 in Rat Serum Using a Hybrid Immuno-Capture LC-MS/MS.AAPS JOURNAL,27(1),13.
MLA Yu, Xiong,et al."Simultaneous Quantification of Total Antibody and Conjugated Payload for DS001 in Rat Serum Using a Hybrid Immuno-Capture LC-MS/MS".AAPS JOURNAL 27.1(2025):13.

入库方式: OAI收割

来源:上海药物研究所

浏览0
下载0
收藏0
其他版本