激光诱导击穿光谱实验数据波动性的影响因素研究

文献类型:期刊论文

作者辛勇; 孙兰香; 丛智博; 杨志家
刊名冶金分析
出版日期2012
卷号32期号:10页码:16-20
关键词激光诱导击穿光谱 等离子体 激光能量 聚焦深度 信号采集位置
ISSN号1000-7571
其他题名An experimental study on the factors causing data fluctuation of laser-induced breakdown spectrometric determination
产权排序1
中文摘要由于激光诱导击穿光谱(LIBS)实验数据容易受各种实验条件的影响,所以采集的光谱数据相对标准偏差较大,导致后续的定量分析精度不高。本文采用激光诱导击穿光谱,以高合金钢样品中Si,Cr,Mn,Fe 4种元素的特征谱线为对象,系统的研究了各种可能引起光谱数据波动的因素,如激光能量及频率、收集位置、焦深等对光谱实验数据波动性的影响。实验发现,激光能量的稳定、合适的聚焦深度等都是提高灵敏度,降低光谱数据相对标准偏差(RSD)的有效手段,而改变信号采集位置及角度则影响不大。
英文摘要Laser-induced breakdown spectrometric(LIBS) determination is susceptible to various experimental conditions,so that the acquired relative standard deviation of its experimental data is large,causing poor precision of the successive qualitative analysis.In this paper,the intensity of the characteristic line of Si,Cr,Mn and Fe was determined by LIBS for the purpose of systematically studying the factors that may cause data fluctuation,such as,laser energy and frequency,acquisition position,and focal depth.It ...
收录类别CSCD
资助信息国家高技术研究发展计划项目(2012AA040608); 国家自然科学基金项目(61004131)
语种中文
CSCD记录号CSCD:4670373
公开日期2012-12-28
源URL[http://ir.sia.ac.cn/handle/173321/10313]  